Share
𝕏 Facebook LinkedIn

Electrodynamics and radiofrequency antenna concepts for human magnetic resonance at 23.5 T (1 GHz) and beyond.

PAPER pubmed Magma (New York, N.Y.) 2016 Engineering / measurement Effect: unclear Evidence: Insufficient

Abstract

OBJECTIVE: This work investigates electrodynamic constraints, explores RF antenna concepts and examines the transmission fields (B 1 (+) ) and RF power deposition of dipole antenna arrays for (1)H magnetic resonance of the human brain at 1 GHz (23.5 T). MATERIALS AND METHODS: Electromagnetic field (EMF) simulations are performed in phantoms with average tissue simulants for dipole antennae using discrete frequencies [300 MHz (7.0 T) to 3 GHz (70.0 T)]. To advance to a human setup EMF simulations are conducted in anatomical human voxel models of the human head using a 20-element dipole array operating at 1 GHz. RESULTS: Our results demonstrate that transmission fields suitable for (1)H MR of the human brain can be achieved at 1 GHz. An increase in transmit channel density around the human head helps to enhance B 1 (+) in the center of the brain. The calculated relative increase in specific absorption rate at 23.5 versus 7.0 T was below 1.4 (in-phase phase setting) and 2.7 (circular polarized phase setting) for the dipole antennae array. CONCLUSION: The benefits of multi-channel dipole antennae at higher frequencies render MR at 23.5 T feasible from an electrodynamic standpoint. This very preliminary finding opens the door on further explorations that might be catalyzed into a 20-T class human MR system.

AI evidence extraction

At a glance
Study type
Engineering / measurement
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF MRI (human magnetic resonance) dipole antenna array · 1000 MHz
Evidence strength
Insufficient
Confidence: 78% · Peer-reviewed: yes

Main findings

EMF simulations in phantoms and anatomical human head voxel models suggest that transmission fields suitable for 1H brain MR can be achieved at 1 GHz (23.5 T). Increasing transmit channel density around the head enhanced B1+ in the brain center. The calculated relative SAR increase at 23.5 T versus 7.0 T was below 1.4 (in-phase) and 2.7 (circular polarized) for the dipole array.

Outcomes measured

  • Transmission fields (B1+)
  • RF power deposition
  • Specific absorption rate (SAR) (relative increase at 23.5 T vs 7.0 T)

Limitations

  • Simulation-based study (EMF simulations in phantoms and voxel models)
  • Described as a "very preliminary" finding
  • SAR results reported as relative increase; absolute SAR not provided in abstract
View raw extracted JSON
{
    "study_type": "engineering",
    "exposure": {
        "band": "RF",
        "source": "MRI (human magnetic resonance) dipole antenna array",
        "frequency_mhz": 1000,
        "sar_wkg": null,
        "duration": null
    },
    "population": null,
    "sample_size": null,
    "outcomes": [
        "Transmission fields (B1+)",
        "RF power deposition",
        "Specific absorption rate (SAR) (relative increase at 23.5 T vs 7.0 T)"
    ],
    "main_findings": "EMF simulations in phantoms and anatomical human head voxel models suggest that transmission fields suitable for 1H brain MR can be achieved at 1 GHz (23.5 T). Increasing transmit channel density around the head enhanced B1+ in the brain center. The calculated relative SAR increase at 23.5 T versus 7.0 T was below 1.4 (in-phase) and 2.7 (circular polarized) for the dipole array.",
    "effect_direction": "unclear",
    "limitations": [
        "Simulation-based study (EMF simulations in phantoms and voxel models)",
        "Described as a \"very preliminary\" finding",
        "SAR results reported as relative increase; absolute SAR not provided in abstract"
    ],
    "evidence_strength": "insufficient",
    "confidence": 0.7800000000000000266453525910037569701671600341796875,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "MRI",
        "ultra-high field",
        "23.5 T",
        "1 GHz",
        "dipole antenna array",
        "B1+",
        "RF power deposition",
        "SAR",
        "EMF simulations",
        "human head voxel model",
        "phantom"
    ],
    "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.