Share
𝕏 Facebook LinkedIn

Radio frequency induced heating of an insulated wire during magnetic resonance imaging.

PAPER pubmed Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference 2016 Engineering / measurement Effect: unclear Evidence: Low

Abstract

Radio frequency (RF) induced power deposition near straight insulated leads excited by different incident electrical field profiles was investigated. 3-D electromagnetic (EM) simulations were used to obtain implant models as well as RF-induced power depositions calculated as the integral of power loss density. For the investigated setups up to 400% discrepancy between the implant model predictions and EM simulation values was observed.

AI evidence extraction

At a glance
Study type
Engineering / measurement
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF magnetic resonance imaging (MRI)
Evidence strength
Low
Confidence: 74% · Peer-reviewed: yes

Main findings

3-D electromagnetic simulations assessed RF-induced power deposition near straight insulated leads under different incident electric field profiles. For the investigated setups, up to a 400% discrepancy was observed between implant model predictions and EM simulation values.

Outcomes measured

  • RF-induced power deposition near insulated leads
  • Discrepancy between implant model predictions and 3-D EM simulation values

Limitations

  • No MRI/RF frequency reported in the abstract
  • No experimental (in vivo/in vitro) validation described; simulations only
  • No quantitative heating/temperature outcomes reported (power deposition only)
  • Sample size and specific lead/implant configurations not detailed

Suggested hubs

  • occupational-exposure (0.2)
    MRI-related RF exposure context, but no specific occupational or human exposure details provided.
View raw extracted JSON
{
    "study_type": "engineering",
    "exposure": {
        "band": "RF",
        "source": "magnetic resonance imaging (MRI)",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": null,
    "sample_size": null,
    "outcomes": [
        "RF-induced power deposition near insulated leads",
        "Discrepancy between implant model predictions and 3-D EM simulation values"
    ],
    "main_findings": "3-D electromagnetic simulations assessed RF-induced power deposition near straight insulated leads under different incident electric field profiles. For the investigated setups, up to a 400% discrepancy was observed between implant model predictions and EM simulation values.",
    "effect_direction": "unclear",
    "limitations": [
        "No MRI/RF frequency reported in the abstract",
        "No experimental (in vivo/in vitro) validation described; simulations only",
        "No quantitative heating/temperature outcomes reported (power deposition only)",
        "Sample size and specific lead/implant configurations not detailed"
    ],
    "evidence_strength": "low",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "radiofrequency",
        "MRI",
        "RF-induced heating",
        "power deposition",
        "insulated wire",
        "leads",
        "electromagnetic simulation",
        "incident electric field"
    ],
    "suggested_hubs": [
        {
            "slug": "occupational-exposure",
            "weight": 0.200000000000000011102230246251565404236316680908203125,
            "reason": "MRI-related RF exposure context, but no specific occupational or human exposure details provided."
        }
    ]
}

AI can be wrong. Always verify against the paper.