Evaluation of MRI RF electromagnetic field induced heating near leads of cochlear implants.
Research
RF Safe Research Library
Jan 1, 2018
This simulation study evaluated RF-induced heating near cochlear implant leads during 1.5 T MRI using a transfer function method. Across studied configurations, the average temperature rise was 0.79 °C and the maximum was 2.80 °C at a whole-body average SAR of 2 W/kg. The authors report that lead trajectory and MRI…
Dependence of RF heating on SAR and implant position in a 1.5T MR system.
Research
RF Safe Research Library
Jan 1, 2007
This phantom study measured RF-induced heating of a stainless steel humerus implant during 1.5T MR imaging while varying SAR, implant angle to B0, and implant position in the RF coil. Temperature rise at implant tips increased linearly with SAR, with larger rises when the implant was parallel to B0. Heating decreased…