Low dose of continuous-wave microwave irradiation did not cause temperature increase in muscles tissue adjacent to titanium alloy implants--an animal study.
Abstract
BACKGROUND: Research studies on the influence of radiofrequency electromagnetic radiation on implants in vitro have failed to investigate temperature changes in the tissues adjacent to the implants under microwave therapy. We therefore, used a rabbit model in an effort to determine the impact of microwave therapy on temperature changes in tissues adjacent to the titanium alloy implants and the safety profile thereof. METHODS: Titanium alloy internal fixation plates were implanted in New Zealand rabbits in the middle of femur. Microwave therapy was performed by a 2450 MHz microwave generator 3 days after the surgery. Temperature changes of muscles adjacent to the implants were recorded under exposure to dose-gradient microwave radiation from 20w to 60w. RESULTS: Significant difference between control and microwave treatment group at peak temperatures (T(peak)) and temperature gap (T(gap) = T(peak)-T(vally)) were observed in deep muscles (T(peak), 41.63 ± 0.21°C vs. 44.40 ± 0.17°C, P < 0.01; T(gap), 5.33 ± 0.21°C vs. 8.10 ± 0.36°C, P < 0.01) and superficial muscles (T(peak), 41.53 ± 0.15°C vs. 42.03 ± 0.23°C, P = 0.04; T(gap), 5.23 ± 0.21°C vs. 5.80 ± 0.17°C, P = 0.013) under 60 w, and deep muscles (T(peak), 40.93 ± 0.25°C vs. 41.87 ± 0.23°C, P = 0.01; T(gap), 4.73 ± 0.20°C vs. 5.63 ± 0.35°C, P = 0.037) under 50w, but not under 20, 30 and 40w. CONCLUSION: Our results suggest that low-dose (20w-40w) continuous-wave microwave irradiation delivered by a 2450 MHz microwave generator might be a promising treatment for patients with titanium alloy internal fixation, as it did not raise temperature in muscle tissues adjacent to the titanium alloy implant.
AI evidence extraction
Main findings
In rabbits with titanium alloy femoral fixation plates, microwave therapy at 2450 MHz produced significant increases in peak temperature and temperature gap in deep and superficial muscles at 60 W, and in deep muscles at 50 W, but not at 20–40 W. The authors conclude that 20–40 W continuous-wave exposure did not raise adjacent muscle temperature.
Outcomes measured
- Temperature changes in muscle tissue adjacent to titanium alloy implants (peak temperature, temperature gap)
Limitations
- Sample size not reported in the abstract
- Exposure duration and other dosimetry details (e.g., SAR, distance, field strength) not reported
- Animal model; generalizability to humans uncertain
- Temperature outcomes only; no clinical endpoints reported
Suggested hubs
-
occupational-exposure
(0.2) Involves microwave exposure, but in a therapeutic/experimental setting rather than occupational; hub match is weak.
View raw extracted JSON
{
"study_type": "animal",
"exposure": {
"band": "microwave",
"source": "microwave therapy",
"frequency_mhz": 2450,
"sar_wkg": null,
"duration": null
},
"population": "New Zealand rabbits with titanium alloy internal fixation plates implanted in the femur",
"sample_size": null,
"outcomes": [
"Temperature changes in muscle tissue adjacent to titanium alloy implants (peak temperature, temperature gap)"
],
"main_findings": "In rabbits with titanium alloy femoral fixation plates, microwave therapy at 2450 MHz produced significant increases in peak temperature and temperature gap in deep and superficial muscles at 60 W, and in deep muscles at 50 W, but not at 20–40 W. The authors conclude that 20–40 W continuous-wave exposure did not raise adjacent muscle temperature.",
"effect_direction": "mixed",
"limitations": [
"Sample size not reported in the abstract",
"Exposure duration and other dosimetry details (e.g., SAR, distance, field strength) not reported",
"Animal model; generalizability to humans uncertain",
"Temperature outcomes only; no clinical endpoints reported"
],
"evidence_strength": "low",
"confidence": 0.7800000000000000266453525910037569701671600341796875,
"peer_reviewed_likely": "yes",
"keywords": [
"microwave therapy",
"2450 MHz",
"continuous-wave",
"titanium alloy implant",
"internal fixation plate",
"rabbit model",
"muscle temperature",
"thermal effects"
],
"suggested_hubs": [
{
"slug": "occupational-exposure",
"weight": 0.200000000000000011102230246251565404236316680908203125,
"reason": "Involves microwave exposure, but in a therapeutic/experimental setting rather than occupational; hub match is weak."
}
]
}
AI can be wrong. Always verify against the paper.
Comments
Log in to comment.
No comments yet.