Preliminary Study of Coagulation Monitoring by Antenna for Treatment during Microwave Coagulation Therapy.

PAPER pubmed The open biomedical engineering journal 2010 Engineering / measurement Effect: unclear Evidence: Insufficient

Abstract

Microwave coagulation therapy (MCT) has been employed mainly for treatment of small size tumors. In the treatment, thin microwave antenna is inserted into the tumor and microwave energy heats up the tumor up to at least 60 degrees C for generation of enough coagulated volume including the target tumor. During the microwave radiation, reflection coefficient of treatment antenna changes significantly. In this paper, possibility of coagulation monitoring was found observing the reflection coefficient change of the antenna by numerical calculations and measurements.

AI evidence extraction

At a glance
Study type
Engineering / measurement
Effect direction
unclear
Population
—
Sample size
—
Exposure
microwave microwave coagulation therapy antenna
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

During microwave radiation in microwave coagulation therapy, the reflection coefficient of the treatment antenna changes significantly. Numerical calculations and measurements suggested the possibility of monitoring coagulation by observing these reflection coefficient changes.

Outcomes measured

  • Change in antenna reflection coefficient during microwave coagulation therapy
  • Feasibility of coagulation monitoring via antenna reflection coefficient

Limitations

  • Frequency and exposure parameters (e.g., power, SAR, duration) not reported in the abstract
  • No sample size or subject/phantom details provided in the abstract
  • Outcome focused on monitoring/measurement feasibility rather than health effects
View raw extracted JSON
{
    "study_type": "engineering",
    "exposure": {
        "band": "microwave",
        "source": "microwave coagulation therapy antenna",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": null,
    "sample_size": null,
    "outcomes": [
        "Change in antenna reflection coefficient during microwave coagulation therapy",
        "Feasibility of coagulation monitoring via antenna reflection coefficient"
    ],
    "main_findings": "During microwave radiation in microwave coagulation therapy, the reflection coefficient of the treatment antenna changes significantly. Numerical calculations and measurements suggested the possibility of monitoring coagulation by observing these reflection coefficient changes.",
    "effect_direction": "unclear",
    "limitations": [
        "Frequency and exposure parameters (e.g., power, SAR, duration) not reported in the abstract",
        "No sample size or subject/phantom details provided in the abstract",
        "Outcome focused on monitoring/measurement feasibility rather than health effects"
    ],
    "evidence_strength": "insufficient",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "microwave coagulation therapy",
        "antenna",
        "reflection coefficient",
        "coagulation monitoring",
        "tumor ablation",
        "numerical calculations",
        "measurements"
    ],
    "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.