Performance of a radio link between a base station and a medical implant utilising the MICS standard.

PAPER pubmed Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference 2004 Engineering / measurement Effect: unclear Evidence: Insufficient

Abstract

Modern medical implants are of increasing complexity and with that, the need for fast and flexible communication with them grows. A wireless system is preferable and an inductive link is the most commonly used. But it has the drawback of a very short range, essentially limited to having the external transceiver touching the patient. The Medical Implant Communication System, MICS, is a standard aimed at improving the communication distance. It operates at a higher frequency band between 402 MHz and 405 MHz. We have by simulations and measurements investigated the channel properties of this band and calculated the link performance for a typical implant application. The result is a link speed between a base station and a bedridden patient of 600 kbit bits per second with a bit error rate of 2% in the downlink to the implant and 1 % in the uplink to the base station. Conclusions on the necessary complexity of the base station are also given.

AI evidence extraction

At a glance
Study type
Engineering / measurement
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF medical implant communication (MICS) base station/implant link
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: unknown

Main findings

Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.

Outcomes measured

  • Channel properties at 402–405 MHz (MICS band)
  • Link speed (kbps)
  • Bit error rate (downlink/uplink)
  • Base station complexity requirements

Limitations

  • No health outcomes assessed (focus is link/channel performance).
  • Sample size and measurement setup details are not provided in the abstract.
  • Exposure metrics such as SAR or field strength are not reported in the abstract.
View raw extracted JSON
{
    "study_type": "engineering",
    "exposure": {
        "band": "RF",
        "source": "medical implant communication (MICS) base station/implant link",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": null,
    "sample_size": null,
    "outcomes": [
        "Channel properties at 402–405 MHz (MICS band)",
        "Link speed (kbps)",
        "Bit error rate (downlink/uplink)",
        "Base station complexity requirements"
    ],
    "main_findings": "Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.",
    "effect_direction": "unclear",
    "limitations": [
        "No health outcomes assessed (focus is link/channel performance).",
        "Sample size and measurement setup details are not provided in the abstract.",
        "Exposure metrics such as SAR or field strength are not reported in the abstract."
    ],
    "evidence_strength": "insufficient",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "unknown",
    "keywords": [
        "medical implant communication system",
        "MICS",
        "402–405 MHz",
        "radio link",
        "base station",
        "implant",
        "channel properties",
        "bit error rate",
        "link speed"
    ],