A new method of radio frequency links by coplanar coils for implantable medical devices.
Abstract
A new method based on coplanar coils for the design of radio frequency links has been developed, to realize the communication between the programming wand and the implantable medical devices with shielding container simply and reliably. With the analysis of electronic and magnetic field theory, the communication model has been established and simulated, and the circuit has been designed and tested. The experimental results are consistent with the simulation fairly well. The voltage transfer ratio of the typical circuit with present parameters can reach as high as 0.02, which can fulfill the requirements of communication.
AI evidence extraction
Main findings
A coplanar-coil based RF link design for communication between a programming wand and implantable medical devices was modeled, simulated, and tested. Experimental results were fairly consistent with simulations, and a typical circuit achieved a voltage transfer ratio up to 0.02, stated as sufficient to meet communication requirements.
Outcomes measured
- Voltage transfer ratio
- Communication link performance between programming wand and implantable medical devices
Limitations
- No RF frequency or exposure parameters reported in the abstract (e.g., frequency, power, SAR).
- No biological/health outcomes assessed; focus is on communication/link performance.
- No sample size or detailed experimental setup described in the abstract.
Suggested hubs
-
engineering
(0.78) Design, modeling, simulation, and testing of an RF communication link for implantable medical devices.
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": "RF",
"source": "implantable medical devices",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"Voltage transfer ratio",
"Communication link performance between programming wand and implantable medical devices"
],
"main_findings": "A coplanar-coil based RF link design for communication between a programming wand and implantable medical devices was modeled, simulated, and tested. Experimental results were fairly consistent with simulations, and a typical circuit achieved a voltage transfer ratio up to 0.02, stated as sufficient to meet communication requirements.",
"effect_direction": "unclear",
"limitations": [
"No RF frequency or exposure parameters reported in the abstract (e.g., frequency, power, SAR).",
"No biological/health outcomes assessed; focus is on communication/link performance.",
"No sample size or detailed experimental setup described in the abstract."
],
"evidence_strength": "insufficient",
"confidence": 0.7399999999999999911182158029987476766109466552734375,
"peer_reviewed_likely": "yes",
"keywords": [
"radio frequency link",
"coplanar coils",
"implantable medical devices",
"programming wand",
"shielding container",
"simulation",
"circuit testing",
"voltage transfer ratio"
],
"suggested_hubs": [
{
"slug": "engineering",
"weight": 0.7800000000000000266453525910037569701671600341796875,
"reason": "Design, modeling, simulation, and testing of an RF communication link for implantable medical devices."
}
]
}
AI can be wrong. Always verify against the paper.