Conductive keratoplasty: a radiofrequency-based technique for the correction of hyperopia.
Abstract
PURPOSE: To evaluate the data on the safety, effectiveness, and stability of conductive keratoplasty (CK), a thermal, radiofrequency-based technique for treating 0.75 to 3.00 diopters (D) of spherical hyperopia. METHODS: A prospective, consecutive series, multicenter clinical trial involving 400 hyperopic eyes was conducted by 19 surgeons at 12 centers. The treatment goal was emmetropia. Cohort follow-up was up to 2 years. RESULTS: At 12 months postoperatively, data were available for 97.5% (354/363) of eyes for efficacy variables and 98% (391/400) of eyes for safety variables. A total of 54% of the CK eyes showed 20/20 or better uncorrected visual acuity, and 92% showed 20/40 or better at 12 months. The mean postoperative manifest refractive spherical equivalent was within 0.50 D in 61% and within 1.00 D in 88%. After CK, eyes were approximately 0.50 D myopic at month 1 and effectively emmetropic at 6 months. At 24 months, there was a 20% loss of initial effect. Safety results showed a 2-line loss of best-corrected visual acuity in 2% of the CK-treated eyes. The incidence of induced cylinder of 2.00 D or greater was below 1%. CONCLUSION: The CK technique corrects low to moderate hyperopia effectively and safely and with acceptable stability. It spares the visual axis, does not require the creation of a large flap, and is not associated with postoperative dry eye. CK has value as an alternative to hyperopic LASIK for patients with hyperopia.
AI evidence extraction
Main findings
In a prospective multicenter clinical trial of 400 hyperopic eyes, 54% achieved 20/20 or better uncorrected visual acuity and 92% achieved 20/40 or better at 12 months. Refractive outcomes were within 0.50 D in 61% and within 1.00 D in 88%; at 24 months there was a reported 20% loss of initial effect. Safety outcomes included a 2-line loss of best-corrected visual acuity in 2% of treated eyes and induced cylinder ≥2.00 D in <1%.
Outcomes measured
- Uncorrected visual acuity (20/20 or better; 20/40 or better)
- Manifest refractive spherical equivalent accuracy (within 0.50 D; within 1.00 D)
- Stability/Regression of refractive effect at 24 months
- Safety: loss of best-corrected visual acuity (2-line loss)
- Safety: induced cylinder (≥2.00 D)
Limitations
- Radiofrequency exposure parameters (e.g., frequency, power, SAR) not reported in abstract
- No comparator/control group described in abstract
- Follow-up reported up to 2 years; longer-term stability beyond 24 months not available in abstract
View raw extracted JSON
{
"study_type": "other",
"exposure": {
"band": "RF",
"source": "medical device (conductive keratoplasty)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": "Hyperopic eyes (0.75 to 3.00 diopters spherical hyperopia) undergoing conductive keratoplasty",
"sample_size": 400,
"outcomes": [
"Uncorrected visual acuity (20/20 or better; 20/40 or better)",
"Manifest refractive spherical equivalent accuracy (within 0.50 D; within 1.00 D)",
"Stability/Regression of refractive effect at 24 months",
"Safety: loss of best-corrected visual acuity (2-line loss)",
"Safety: induced cylinder (≥2.00 D)"
],
"main_findings": "In a prospective multicenter clinical trial of 400 hyperopic eyes, 54% achieved 20/20 or better uncorrected visual acuity and 92% achieved 20/40 or better at 12 months. Refractive outcomes were within 0.50 D in 61% and within 1.00 D in 88%; at 24 months there was a reported 20% loss of initial effect. Safety outcomes included a 2-line loss of best-corrected visual acuity in 2% of treated eyes and induced cylinder ≥2.00 D in <1%.",
"effect_direction": "benefit",
"limitations": [
"Radiofrequency exposure parameters (e.g., frequency, power, SAR) not reported in abstract",
"No comparator/control group described in abstract",
"Follow-up reported up to 2 years; longer-term stability beyond 24 months not available in abstract"
],
"evidence_strength": "moderate",
"confidence": 0.7399999999999999911182158029987476766109466552734375,
"peer_reviewed_likely": "yes",