Determination of blood longitudinal relaxation time (T1) at high magnetic field strengths.

PAPER pubmed Magnetic resonance imaging 2007 Other Effect: unclear Evidence: Insufficient

Abstract

In this study, a circulation system was used to measure T(1) values of bovine blood under physiological conditions at field strengths of 4.7, 7 and 9.4 T. Results show that T(1) increases linearly with magnetic field B(0) and can be described with the equation T(1)=129 ms/T B(0)+1167 ms for magnetic field strengths between 1.5 and 9.4 T.

AI evidence extraction

At a glance
Study type
Other
Effect direction
unclear
Population
Bovine blood (circulating system)
Sample size
—
Exposure
MRI (static magnetic field B0)
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

Using a circulation system under physiological conditions, bovine blood T1 increased linearly with magnetic field strength (B0) across 4.7, 7, and 9.4 T. The relationship was described as T1 = 129 ms/T * B0 + 1167 ms for magnetic field strengths between 1.5 and 9.4 T.

Outcomes measured

  • Blood longitudinal relaxation time (T1)

Limitations

  • Sample size not reported in abstract
  • Study uses bovine blood in a circulation system; generalizability to humans not stated
  • No health outcomes assessed; outcome is a measurement parameter (T1)
View raw extracted JSON
{
    "study_type": "other",
    "exposure": {
        "band": null,
        "source": "MRI (static magnetic field B0)",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": "Bovine blood (circulating system)",
    "sample_size": null,
    "outcomes": [
        "Blood longitudinal relaxation time (T1)"
    ],
    "main_findings": "Using a circulation system under physiological conditions, bovine blood T1 increased linearly with magnetic field strength (B0) across 4.7, 7, and 9.4 T. The relationship was described as T1 = 129 ms/T * B0 + 1167 ms for magnetic field strengths between 1.5 and 9.4 T.",
    "effect_direction": "unclear",
    "limitations": [
        "Sample size not reported in abstract",
        "Study uses bovine blood in a circulation system; generalizability to humans not stated",
        "No health outcomes assessed; outcome is a measurement parameter (T1)"
    ],
    "evidence_strength": "insufficient",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "MRI",
        "static magnetic field",
        "B0",
        "T1 relaxation time",
        "blood",
        "high field strength",
        "4.7 T",
        "7 T",
        "9.4 T"
    ],
    "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.