Isolated Atrial Lead Conduction Delay following Right Atrial Radiofrequency Maze Procedure.

PAPER pubmed ISRN cardiology 2011 Case report Effect: harm Evidence: Very low

Abstract

A 60-year-old man with a dual-chamber implantable cardioverter defibrillator and severe dyspnea on exertion due to apical hypertrophic cardiomyopathy underwent a septal myectomy and radiofrequency maze procedure. Following the procedure a persistent delay in atrial sensing was observed and was most likely a result of iatrogenic conduction delay from right atrial ablation lines. These observations suggest that atrial conduction properties can be altered during the surgical maze procedure and should be considered in the differential diagnosis of sensing or pacing malfunction.

AI evidence extraction

At a glance
Study type
Case report
Effect direction
harm
Population
60-year-old man with a dual-chamber implantable cardioverter defibrillator and apical hypertrophic cardiomyopathy
Sample size
1
Exposure
radiofrequency maze procedure (right atrial ablation lines)
Evidence strength
Very low
Confidence: 74% · Peer-reviewed: yes

Main findings

After septal myectomy and a radiofrequency maze procedure, a persistent delay in atrial sensing was observed, most likely due to iatrogenic conduction delay from right atrial ablation lines. The report suggests atrial conduction properties can be altered during surgical maze procedures and may contribute to apparent sensing/pacing malfunction.

Outcomes measured

  • atrial sensing delay
  • iatrogenic atrial conduction delay
  • sensing or pacing malfunction considerations

Limitations

  • Single-patient case report
  • No radiofrequency exposure parameters (e.g., frequency, power, duration) reported
  • Causality inferred as most likely; no confirmatory testing described in abstract
View raw extracted JSON
{
    "study_type": "case_report",
    "exposure": {
        "band": null,
        "source": "radiofrequency maze procedure (right atrial ablation lines)",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": "60-year-old man with a dual-chamber implantable cardioverter defibrillator and apical hypertrophic cardiomyopathy",
    "sample_size": 1,
    "outcomes": [
        "atrial sensing delay",
        "iatrogenic atrial conduction delay",
        "sensing or pacing malfunction considerations"
    ],
    "main_findings": "After septal myectomy and a radiofrequency maze procedure, a persistent delay in atrial sensing was observed, most likely due to iatrogenic conduction delay from right atrial ablation lines. The report suggests atrial conduction properties can be altered during surgical maze procedures and may contribute to apparent sensing/pacing malfunction.",
    "effect_direction": "harm",
    "limitations": [
        "Single-patient case report",
        "No radiofrequency exposure parameters (e.g., frequency, power, duration) reported",
        "Causality inferred as most likely; no confirmatory testing described in abstract"
    ],
    "evidence_strength": "very_low",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "radiofrequency maze",
        "right atrial ablation",
        "atrial lead",
        "conduction delay",
        "atrial sensing",
        "implantable cardioverter defibrillator",
        "hypertrophic cardiomyopathy"
    ],
    "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.