Ultrafast volumetric B1 (+) mapping for improved radiofrequency shimming in 3 tesla body MRI.

PAPER pubmed Journal of magnetic resonance imaging : JMRI 2014 Engineering / measurement Effect: benefit Evidence: Low

Abstract

PURPOSE: To evaluate the use of the recently proposed ultrafast B1 (+) mapping approach DREAM (Dual Refocusing Echo Acquisition Mode) for a refinement of patient adaptive radiofrequency (RF) shimming. MATERIALS AND METHODS: Volumetric DREAM B1 (+) calibration scans centered in the upper abdomen were acquired in 20 patients and three volunteers with written informed consent at a clinical dual source 3 Tesla (T) MR system. Based on these data, RF transmit settings were optimized by central-slice based RF-shimming (CS-RF shim) and by a refined, multi-slice adaptive approach (MS-RF shim). Simulations were performed to compare flip angle accuracy and B1 (+) homogeneity (cv = stddev/mean) achieved by CS-RF shim versus MS-RF shim for transversal and coronal slices, and for volume shimming on the spine. RESULTS: By MS-RF shim, mean deviation from nominal flip angle was reduced to less than 11% in all slices, all targets, and all subjects. Relative improvements in B1 (+) cv (MS-RF shim versus CS-RF) were up to 14%/39%/47% in transversal slices/coronal slices/ spine area. CONCLUSION: Volumetric information about B1 (+) can be used to further improve the accuracy and homogeneity of the B1 (+) field yielding higher diagnostic confidence, and will also be of value for various quantitative methods which are sensitive to flip angle imperfections.

AI evidence extraction

At a glance
Study type
Engineering / measurement
Effect direction
benefit
Population
20 patients and three volunteers
Sample size
23
Exposure
RF MRI
Evidence strength
Low
Confidence: 74% · Peer-reviewed: yes

Main findings

Using multi-slice adaptive RF shimming (MS-RF shim) based on volumetric DREAM B1+ calibration scans, mean deviation from nominal flip angle was reduced to less than 11% across slices/targets/subjects. Relative improvements in B1+ homogeneity (cv) for MS-RF shim versus central-slice RF shimming were reported up to 14% (transversal), 39% (coronal), and 47% (spine area).

Outcomes measured

  • B1+ mapping performance (DREAM)
  • Flip angle accuracy (deviation from nominal)
  • B1+ homogeneity (coefficient of variation, cv)
  • Performance of RF shimming approaches (CS-RF shim vs MS-RF shim)

Suggested hubs

  • mri-rf (0.9)
    Study evaluates RF (B1+) mapping and RF shimming methods in a 3T clinical MRI system.
View raw extracted JSON
{
    "study_type": "engineering",
    "exposure": {
        "band": "RF",
        "source": "MRI",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": null
    },
    "population": "20 patients and three volunteers",
    "sample_size": 23,
    "outcomes": [
        "B1+ mapping performance (DREAM)",
        "Flip angle accuracy (deviation from nominal)",
        "B1+ homogeneity (coefficient of variation, cv)",
        "Performance of RF shimming approaches (CS-RF shim vs MS-RF shim)"
    ],
    "main_findings": "Using multi-slice adaptive RF shimming (MS-RF shim) based on volumetric DREAM B1+ calibration scans, mean deviation from nominal flip angle was reduced to less than 11% across slices/targets/subjects. Relative improvements in B1+ homogeneity (cv) for MS-RF shim versus central-slice RF shimming were reported up to 14% (transversal), 39% (coronal), and 47% (spine area).",
    "effect_direction": "benefit",
    "limitations": [],
    "evidence_strength": "low",
    "confidence": 0.7399999999999999911182158029987476766109466552734375,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "B1+ mapping",
        "DREAM",
        "Dual Refocusing Echo Acquisition Mode",
        "radiofrequency shimming",
        "RF transmit settings",
        "3 Tesla",
        "body MRI",
        "flip angle",
        "B1+ homogeneity",
        "upper abdomen"
    ],
    "suggested_hubs": [
        {
            "slug": "mri-rf",
            "weight": 0.90000000000000002220446049250313080847263336181640625,
            "reason": "Study evaluates RF (B1+) mapping and RF shimming methods in a 3T clinical MRI system."
        }
    ]
}