Environmental radiofrequency electromagnetic field levels in a department of pediatrics
Abstract
Environmental radiofrequency electromagnetic field levels in a department of pediatrics Besset D, Selmaoui B, Tourneux P, Leke A, Delanaud S, de Seze R, Stephan Blanchard E. Environmental radiofrequency electromagnetic field levels in a department of pediatrics. Environ Res. 2019 Nov 6:108894. doi: 10.1016/j.envres.2019.108894. Abstract Preterm neonates constitute a vulnerable population that is highly sensitive to its environment. Given the increased use of wireless communication devices (mobile and digital enhanced cordless telecommunications, WiFi networks, etc.), neonates hospitalized in a department of pediatrics are potentially exposed to radiofrequency electromagnetic fields (RF-EMF). Strikingly, data on RF-EMF levels in pediatric units have not previously been published. The objective of the present study was thus to quantify the RF-EMF levels in a 34-bed tertiary department of pediatrics with a neonatal critical care unit (NCCU) and a neonatal intensive care unit (NICU). To this end, we used triaxle antenna dosimeters to map the RF-EMF levels in the environment and to measure spot emissions from medical devices. In a first set of experiments, RF-EMF levels at 144 points in the staff area and in the children's rooms in the NCCU and NICU were evaluated over a 24-h period. In a second set of measurements performed in a Faraday chamber, we measured the RF-EMF levels emitted by the medical devices to which neonates are potentially exposed in the department of pediatrics. The RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05). Although the two units did not differ significantly with regard to the average maximum values, the single greatest value recorded in the NCCU (6 V/m GSM + UMTS 900 (UL) frequency band, in the staff area) was more than twice that recorded in the NICU (3.70 V/m in the UMTS 2100 (UL) frequency band, in the children's rooms). The NCCU and NICU did not differ significantly with regard to the time during which the RF-EMF level at each measurement point was more than two standard deviations above its mean. The RF-EMF level was significantly higher during the day than during the night (p < 0.001). The various medical devices used in the NICU did not emit detectable amounts of RF. Overall, RF-EMF levels in the NCCU and NICU were very low. It is probable that the RF-EMFs measured here were primarily generated by the parents' and staff members' activities, rather than by medical devices. However, a combination of low-level, chronic exposure with transient, elevated peak values in a vulnerable population of preterm neonates may be of particular concern. In a department of pediatrics, decreasing preterm neonates' exposure to RF-EMFs should primarily involve a limitation on the use of wireless communication devices by staff members and parents. ncbi.nlm.nih.gov
AI evidence extraction
Main findings
RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05), and higher during the day than during the night (p < 0.001). The greatest recorded value was 6 V/m (GSM+UMTS 900 UL) in the NCCU staff area versus 3.70 V/m (UMTS 2100 UL) in NICU children's rooms; medical devices tested in the NICU did not emit detectable RF. Overall, measured RF-EMF levels in both units were described as very low and were attributed primarily to parents’ and staff members’ wireless device use rather than medical devices.
Outcomes measured
- Environmental RF-EMF levels (V/m) in NCCU and NICU
- Peak and average maximum RF-EMF values by unit and location
- Day vs night RF-EMF levels
- Detectable RF emissions from medical devices (Faraday chamber)
Limitations
- Single-site study in one 34-bed tertiary pediatrics department
- Frequency bands and metrics beyond those explicitly reported are not detailed in the abstract
- No health outcomes assessed; exposure characterization only
Suggested hubs
-
school-wi-fi
(0.2) WiFi networks mentioned as a potential source, though setting is a hospital neonatal unit rather than a school.
-
occupational-exposure
(0.35) Measurements include staff areas and suggest staff device use as a primary RF-EMF source.
View raw extracted JSON
{
"study_type": "exposure_assessment",
"exposure": {
"band": "RF",
"source": "wireless communication devices (mobile, DECT, WiFi) and medical devices (assessed)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": "24-h measurements at 144 points; additional spot measurements in Faraday chamber"
},
"population": "Preterm neonates in a tertiary department of pediatrics (NCCU and NICU); environmental measurements in staff areas and children's rooms",
"sample_size": null,
"outcomes": [
"Environmental RF-EMF levels (V/m) in NCCU and NICU",
"Peak and average maximum RF-EMF values by unit and location",
"Day vs night RF-EMF levels",
"Detectable RF emissions from medical devices (Faraday chamber)"
],
"main_findings": "RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05), and higher during the day than during the night (p < 0.001). The greatest recorded value was 6 V/m (GSM+UMTS 900 UL) in the NCCU staff area versus 3.70 V/m (UMTS 2100 UL) in NICU children's rooms; medical devices tested in the NICU did not emit detectable RF. Overall, measured RF-EMF levels in both units were described as very low and were attributed primarily to parents’ and staff members’ wireless device use rather than medical devices.",
"effect_direction": "unclear",
"limitations": [
"Single-site study in one 34-bed tertiary pediatrics department",
"Frequency bands and metrics beyond those explicitly reported are not detailed in the abstract",
"No health outcomes assessed; exposure characterization only"
],
"evidence_strength": "insufficient",
"confidence": 0.7800000000000000266453525910037569701671600341796875,
"peer_reviewed_likely": "yes",
"keywords": [
"RF-EMF",
"pediatrics",
"neonatal critical care unit",
"neonatal intensive care unit",
"preterm neonates",
"environmental measurements",
"dosimeters",
"GSM",
"UMTS",
"WiFi",
"DECT",
"Faraday chamber"
],
"suggested_hubs": [
{
"slug": "school-wi-fi",
"weight": 0.200000000000000011102230246251565404236316680908203125,
"reason": "WiFi networks mentioned as a potential source, though setting is a hospital neonatal unit rather than a school."
},
{
"slug": "occupational-exposure",
"weight": 0.34999999999999997779553950749686919152736663818359375,
"reason": "Measurements include staff areas and suggest staff device use as a primary RF-EMF source."
}
]
}