Environmental radiofrequency electromagnetic field levels in a department of pediatrics

PAPER manual 2019 Exposure assessment Effect: unclear Evidence: Insufficient

Abstract

Environmental radiofrequency electromagnetic field levels in a department of pediatrics Besset D, Selmaoui B, Tourneux P, Leke A, Delanaud S, de Seze R, Stephan Blanchard E. Environmental radiofrequency electromagnetic field levels in a department of pediatrics. Environ Res. 2019 Nov 6:108894. doi: 10.1016/j.envres.2019.108894. Abstract Preterm neonates constitute a vulnerable population that is highly sensitive to its environment. Given the increased use of wireless communication devices (mobile and digital enhanced cordless telecommunications, WiFi networks, etc.), neonates hospitalized in a department of pediatrics are potentially exposed to radiofrequency electromagnetic fields (RF-EMF). Strikingly, data on RF-EMF levels in pediatric units have not previously been published. The objective of the present study was thus to quantify the RF-EMF levels in a 34-bed tertiary department of pediatrics with a neonatal critical care unit (NCCU) and a neonatal intensive care unit (NICU). To this end, we used triaxle antenna dosimeters to map the RF-EMF levels in the environment and to measure spot emissions from medical devices. In a first set of experiments, RF-EMF levels at 144 points in the staff area and in the children's rooms in the NCCU and NICU were evaluated over a 24-h period. In a second set of measurements performed in a Faraday chamber, we measured the RF-EMF levels emitted by the medical devices to which neonates are potentially exposed in the department of pediatrics. The RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05). Although the two units did not differ significantly with regard to the average maximum values, the single greatest value recorded in the NCCU (6 V/m GSM + UMTS 900 (UL) frequency band, in the staff area) was more than twice that recorded in the NICU (3.70 V/m in the UMTS 2100 (UL) frequency band, in the children's rooms). The NCCU and NICU did not differ significantly with regard to the time during which the RF-EMF level at each measurement point was more than two standard deviations above its mean. The RF-EMF level was significantly higher during the day than during the night (p < 0.001). The various medical devices used in the NICU did not emit detectable amounts of RF. Overall, RF-EMF levels in the NCCU and NICU were very low. It is probable that the RF-EMFs measured here were primarily generated by the parents' and staff members' activities, rather than by medical devices. However, a combination of low-level, chronic exposure with transient, elevated peak values in a vulnerable population of preterm neonates may be of particular concern. In a department of pediatrics, decreasing preterm neonates' exposure to RF-EMFs should primarily involve a limitation on the use of wireless communication devices by staff members and parents. ncbi.nlm.nih.gov

AI evidence extraction

At a glance
Study type
Exposure assessment
Effect direction
unclear
Population
Preterm neonates in a tertiary department of pediatrics (NCCU and NICU); environmental measurements in staff areas and children's rooms
Sample size
—
Exposure
RF wireless communication devices (mobile, DECT, WiFi) and medical devices (assessed) · 24-h measurements at 144 points; additional spot measurements in Faraday chamber
Evidence strength
Insufficient
Confidence: 78% · Peer-reviewed: yes

Main findings

RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05), and higher during the day than during the night (p < 0.001). The greatest recorded value was 6 V/m (GSM+UMTS 900 UL) in the NCCU staff area versus 3.70 V/m (UMTS 2100 UL) in NICU children's rooms; medical devices tested in the NICU did not emit detectable RF. Overall, measured RF-EMF levels in both units were described as very low and were attributed primarily to parents’ and staff members’ wireless device use rather than medical devices.

Outcomes measured

  • Environmental RF-EMF levels (V/m) in NCCU and NICU
  • Peak and average maximum RF-EMF values by unit and location
  • Day vs night RF-EMF levels
  • Detectable RF emissions from medical devices (Faraday chamber)

Limitations

  • Single-site study in one 34-bed tertiary pediatrics department
  • Frequency bands and metrics beyond those explicitly reported are not detailed in the abstract
  • No health outcomes assessed; exposure characterization only

Suggested hubs

  • school-wi-fi (0.2)
    WiFi networks mentioned as a potential source, though setting is a hospital neonatal unit rather than a school.
  • occupational-exposure (0.35)
    Measurements include staff areas and suggest staff device use as a primary RF-EMF source.
View raw extracted JSON
{
    "study_type": "exposure_assessment",
    "exposure": {
        "band": "RF",
        "source": "wireless communication devices (mobile, DECT, WiFi) and medical devices (assessed)",
        "frequency_mhz": null,
        "sar_wkg": null,
        "duration": "24-h measurements at 144 points; additional spot measurements in Faraday chamber"
    },
    "population": "Preterm neonates in a tertiary department of pediatrics (NCCU and NICU); environmental measurements in staff areas and children's rooms",
    "sample_size": null,
    "outcomes": [
        "Environmental RF-EMF levels (V/m) in NCCU and NICU",
        "Peak and average maximum RF-EMF values by unit and location",
        "Day vs night RF-EMF levels",
        "Detectable RF emissions from medical devices (Faraday chamber)"
    ],
    "main_findings": "RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05), and higher during the day than during the night (p < 0.001). The greatest recorded value was 6 V/m (GSM+UMTS 900 UL) in the NCCU staff area versus 3.70 V/m (UMTS 2100 UL) in NICU children's rooms; medical devices tested in the NICU did not emit detectable RF. Overall, measured RF-EMF levels in both units were described as very low and were attributed primarily to parents’ and staff members’ wireless device use rather than medical devices.",
    "effect_direction": "unclear",
    "limitations": [
        "Single-site study in one 34-bed tertiary pediatrics department",
        "Frequency bands and metrics beyond those explicitly reported are not detailed in the abstract",
        "No health outcomes assessed; exposure characterization only"
    ],
    "evidence_strength": "insufficient",
    "confidence": 0.7800000000000000266453525910037569701671600341796875,
    "peer_reviewed_likely": "yes",
    "keywords": [
        "RF-EMF",
        "pediatrics",
        "neonatal critical care unit",
        "neonatal intensive care unit",
        "preterm neonates",
        "environmental measurements",
        "dosimeters",
        "GSM",
        "UMTS",
        "WiFi",
        "DECT",
        "Faraday chamber"
    ],
    "suggested_hubs": [
        {
            "slug": "school-wi-fi",
            "weight": 0.200000000000000011102230246251565404236316680908203125,
            "reason": "WiFi networks mentioned as a potential source, though setting is a hospital neonatal unit rather than a school."
        },
        {
            "slug": "occupational-exposure",
            "weight": 0.34999999999999997779553950749686919152736663818359375,
            "reason": "Measurements include staff areas and suggest staff device use as a primary RF-EMF source."
        }
    ]
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.