Wideband Self-Grounded Bow-Tie Antenna for Thermal MR.
Research
RF Safe Research Library
Jan 1, 2020
This engineering study designed and tested a compact wideband self-grounded bow-tie RF antenna module for a whole-body 7.0 T MRI system, aiming to enable 1H/19F MRI, MR thermometry, and broadband RF thermal intervention. The authors report validated B1 efficiency measurements and phantom-based heating experiments…
Effects of Anatomical Differences on Electromagnetic Fields, SAR, and Temperature Change.
Research
RF Safe Research Library
Jan 1, 2016
This engineering study evaluates how anatomical/geometric differences affect MRI RF safety metrics. Simulations and a bovine-phantom experiment found that similar transmit B-field patterns can coincide with notably different peak SAR and temperature rise. The authors highlight challenges in using…