Radiofrequency applicator concepts for thermal magnetic resonance of brain tumors at 297 MHz (7.0 Tesla).
Research
RF Safe Research Library
Jan 1, 2020
This engineering/simulation study evaluates RF applicator concepts for ThermalMR brain tumor heating at 297 MHz (7.0 Tesla) using clinically realistic GBM-based models. EMF simulations compared 10 applicator designs and two SAR-based planning algorithms for small and large target volumes. Power-focused optimization…