Archive

3 posts

Showing results for: 7.0 T

Advanced Radio Frequency Applicators for Thermal Magnetic Resonance Theranostics of Brain Tumors.

Research RF Safe Research Library Jan 1, 2023

This engineering/simulation study evaluates RF applicator array designs for Thermal Magnetic Resonance (ThermalMR), which combines MRI with RF hyperthermia for brain tumor theranostics. Using EMF and temperature simulations at 7.0 T, 9.4 T, and 10.5 T, hybrid loop+SGBT dipole arrays are reported to provide superior…

Radiofrequency applicator concepts for thermal magnetic resonance of brain tumors at 297 MHz (7.0 Tesla).

Research RF Safe Research Library Jan 1, 2020

This engineering/simulation study evaluates RF applicator concepts for ThermalMR brain tumor heating at 297 MHz (7.0 Tesla) using clinically realistic GBM-based models. EMF simulations compared 10 applicator designs and two SAR-based planning algorithms for small and large target volumes. Power-focused optimization…

Wideband Self-Grounded Bow-Tie Antenna for Thermal MR.

Research RF Safe Research Library Jan 1, 2020

This engineering study designed and tested a compact wideband self-grounded bow-tie RF antenna module for a whole-body 7.0 T MRI system, aiming to enable 1H/19F MRI, MR thermometry, and broadband RF thermal intervention. The authors report validated B1 efficiency measurements and phantom-based heating experiments…

Page 1 / 1